10 matches found
CVE-2020-12040
CVE-2020-12040 affects Sigma Spectrum Infusion System v6.x (model 35700BAX) and Baxter Spectrum Infusion System v8.x (model 35700BAX2); root cause is unauthenticated clear-text transmission of system status and operational data at the application layer, enabling eavesdropping or MiTM if network c...
CVE-2020-12039
CVE-2020-12039 affects Baxter Sigma Spectrum Infusion System v6.x (35700BAX) and v8.x (35700BAX2); hard-coded passwords entered via keypad grant access to biomedical menus, including device settings, calibration values, and WBM network configuration. This is a local/physical-access issue with pub...
CVE-2020-12047
CVE-2020-12047 affects Baxter Spectrum WBM when used with Spectrum v8.x (model 35700BAX2) in factory-default wireless config; the WBM enables an FTP service with hard-coded credentials (versions v17, v20D29–v20D32, v22D24). Root cause: hard-coded credentials enabling FTP access. Impact per source...
CVE-2014-5433
CVE-2014-5433 affects Baxter’s SIGMA Spectrum Infusion System (Version 6.05, model 35700BAX) with the Wireless Battery Module (WBM) Version 16. The connected disclosures describe multiple issues: a hard-coded password enabling unauthorized management access (CWE-259), an authentication bypass tie...
CVE-2020-12041
The Baxter Spectrum Wireless Battery Module (WBM) telnet CLI vulnerability CVE-2020-12041 affects Baxter Spectrum WBM components (v17, v20D29–D32, v22D24). The Telnet interface grants access to sensitive data and permits temporary changes to network settings, including rebooting the WBM, with cha...
CVE-2014-5431
CVE-2014-5431 affects Baxter SIGMA Spectrum Infusion System 6.05 with Wireless Battery Module 16. The issue is a hard-coded password that grants access to WBM management functions, enabling unauthorized configuration changes (e.g., wireless status and phase-complete audible alarm). The vulnerabil...
CVE-2020-12043
CVE-2020-12043 affects Baxter Spectrum WBM wireless configurations (versions v17, v20D29–D32, v22D24). The FTP service on the WBM remains accessible until the unit is rebooted, due to an operation on a resource after expiration/release. NVD lists CVSSv3.1 base score 9.8; ICS advisory notes CVSSv3...
CVE-2020-12045
CVE-2020-12045 affects Baxter Spectrum WBM when used with Baxter Spectrum v8.x; the WBM runs a Telnet service on port 1023 with hard-coded credentials. Connected sources document a Telnet exposure tied to WBM/Spectrum configurations and assign high/severe CVSS values (up to 9.8) for this vulnerab...
CVE-2014-5432
CVE-2014-5432 affects Baxter SIGMA Spectrum Infusion System v6.05 (model 35700BAX) with Wireless Battery Module (WBM) v16. The root cause includes unauthenticated remote SSH access (Port 22) and related credential exposure. Consequences: remote attacker could modify WBM configuration and retrieve...
CVE-2014-5434
The CVE-2014-5434 entry concerns Baxter’s SIGMA Spectrum Infusion System: version 6.05 (model 35700BAX) with Wireless Battery Module (WBM) version 16, which contains a default, hard-coded credential used with FTP. The vulnerability allows remote access in some vectors (three vulnerabilities remot...